SC Series Instruction Manual     page 81
AC
/A
C-L
VDT I
nput Ch
annel
11
11.3  Calibration Procedure
If you are not familiar with operating the instrument in the SETUP menu mode,
see “SETUP Menu mode” on page 23.  A listing of all menu items is given in
“Setup Menu Reference” on page 157.
Step 1: Wire the transducer to the channel’s connector.
Step 2: Enter the 
CALIBRATION TYPE
.  
You must have the capability to apply either two, three, or five known dis-
placements to the transducer.  The 
CALIBRATION TYPE
 menu item allows you to
specify how many known displacements will be applied during calibration.
Step 3: Enter the
 CALIBRATION DATA.
Consult the Certificate of Calibration for the transducer when entering infor-
mation in the 
CALIBRATION DATA
 sub-menu.
Step 4: Determine the LVDT’s electrical null point, then physically mount it.
The electrical null is the armature position that produces the least electrical
output.  Once this point has been determined, the transducer is physically
clamped into position.  See “Electrical Null and Transducer Mounting” on
page 93.
Step 5: Perform the calibration.
Using the 
CALIBRATE
 menu item starts the calibration process.  You will be
prompted to apply displacements to the transducer as required.