 page 98      008-0608-00
12.2  Wiring
12.2.1 Channel 
Connector
Connect the amplified transducer, in-line amplifier or DC-DC LVDT to a High-Level
Input channel by wiring it to the 12-pin connector of that channel.  The Customer Infor-
mation Sheet that shipped with the instrument describes which cards are installed in
each channel. The pin-out for this connector is shown on the following table.
The Analog Output and Analog Return pins are electrically isolated from all other pins
on the instrument. 
Table 1-6: High-Level Input Channel Pin Connections
Pin
Label
Function
Reference Pin
1 (top)
+EXC
(+)Excitation
10
2
SHUNT1
Shunt Cal Relay
3
3
-SHUNT2
Shunt Cal Relay
2
4
-EXC
(-)Excitation
10
5
+SIG
(+)Signal
10
6
-SIG
(-)Signal
10
7
+OUT
Analog Output
8
8
-OUT
Analog Return
-
9
N/C
No Connection
-
10
DGND
Digital Ground
-
11
AUX1
Auxiliary Function 1
 (connect to pin 10 to activate)
10
12
(bottom)
AUX2
Auxiliary Function 2
 (connect to pin 10 to activate)
10
CAUTION
Identify the amplified transducer or in-line amplifier according to the Option 
Code or model name as shown on its serial number tag.  Use this information 
to select the correct wiring diagram on the following pages. Incorrect wiring can 
damage both the transducer and the instrument.
Failure to comply with these instructions may result in product damage.