Pulse Period
Function Group
Menu Map
Description
Programming Codes
See Also
(MOD)
4
This 
 lets you set a value for the internal pulse generator’s
pulse period. The range of acceptable values is from 2 
 to 65.5
ms. The factory preset value is 2 ms. When this feature is active, its
current value is displayed in the active entry area.
SCPI: 
PULSe:TIMing:PERiod <num>[time suffix] or
Analyzer: 
NONE
 Pulse Menu
Pulse Rate
Function Group
Menu Map
Description
Programming Codes
See Also
MODULATION
4
This 
 lets you set the internal pulse generators pulse repetition
rate. The repetition rate can vary from 15.26 Hz to 500 
 The
factory preset value is 500 Hz. When this feature is active, its current
value is displayed in the active entry area.
SCPI: 
PULse:FREQuency <num>[freq suffix] or
Analyzer: 
NONE
 Pulse Menu
 Operating and Programming Reference HP 8360