Select Power Sweep (asterisk on).
Press (SINGLE].
The synthesizer performs a power sweep beginning at -20 
 and
ending at 
 
 The power meter indicates 
 
Power Slope Operation
This function allows for compensation of high frequency system or
cable losses by linearly increasing the power output as the frequency
increases. For this example refer to the “Menu Map” section.
Press 
Power Slope 
, the active entry area displays:
 RF SLOPE:
X. XX 
 
where X is a numeric value.
Power slope is now active, notice that an asterisk is next to the key
label.
Use the entry keys, rotary knob, or arrow keys to enter a value for
the linear slope.
Press Power Slope again to turn this feature off.