Справочник Пользователя для Barco K5960008-03

Скачать
Страница из 61
52 __________________________________________________________Barco - DCMS - User Manual
Verification
9
Verification
The verification table below lists an example output of a verification table. After
the calibration process, the user can start a verification of the results by pressing
the ‘Start verification’ button in the calibration dialog. The layout of this table is
always the same but the content may depend on the type of display that is
calibrated. The example below is taken after calibration of a Barco MDPW476.
 Table 2: Verification table
Verification is done by measuring the color and luminance of a set of gray levels,
including white. The measured results are compared to the calculated theoretical
values and printed on screen.