Inventory rack 
IR-216U
Inventory rack 
IR-220U
Rack Capacity
IR-216U
IR-220U
IR-312UN
For 2-inch box
For 3-inch box
6
6
12
MDF-U33V
Distributed by:
©2008 SANYO Printed in Japan 2008.12 MA
SHR128
Dimensions
[Unit : mm  (inch)]
Specifications
Model
Temperature range
Maximum cooling 
performance
Exterior dimensions
(W x D x H)
Interior dimensions
(W x D x H)
Net weight
Effective capacity
Shelves
Access port
Compressor
Refrigerant
Alarm
Remote alarm contact
Accessories
MDF-U33V
–50˚C to –86˚C (1˚C increments)
–86˚C 
(Ambient temp. 30˚C)
670 x 867 x 1860mm
(26.4” x 34.1” x 73.2”)
490 x 600 x 1140mm
(19.3” x 23.6” x 44.9”)
Approx. 255kg (562 lbs.)
333 L (11.8 cu.ft.)
Stainless steel, 3 shelves 
17mm diameter, 3 locations 
(back, bottom left/right corner)
Hermetic type, Output: 450 W 
(high stage side), 750 W (low 
stage side)
HFC
High/low temperature, Power 
failure, Filter check, Self 
diagnostics, Door check, Remote 
alarm contact, Part replacement 
notification
Allowable contact capacity: 
DC 30V, 2A
1 set of keys, 1 scraper
Features
Pull-down characteristics
Pull-up characteristics during power failure
CO
2
 backup system operation characteristics
Performance Data
SAFETY
• Alarm lamp and buzzer offer secure 
warning of power failure or abnormal 
temperature increase.
• High and low temperature warning 
provides an audible and visual alarm 
when the temperature deviates more 
than 
±5˚C to ±20˚C (adjustable) from the 
set point.
• Alarm ring-back function ensures that 
buzzer will resume operation, should 
alarm conditions continue after it is 
silenced.
• Microprocessor-controlled filter-clog 
check function protects the refrigeration 
circuit.
• The rugged, one-handed 
outer door latch has a hole to 
allow a padlock to securely 
protect valuable samples.
• Control panel with digital 
display for easy operation.
Options
Circular recorder 
MTR-G85
LCO
2
 backup system
CVK-UB2
LN
2
 backup system
CVK-UBN2
SANYO DAQ 
(Data Acquisition) system 
MTR-480/MTR-2000
*Cooling performance is indicated by the temperature reached at the 
center of the freezer (at ambient temperature of 30
°C with no load). 
In order to use the freezer at a stable temperature for a long time, it is 
recommended that the temperature be set to at least 5
°C higher than 
the indicated lowest temperature. 
In addition, depending on the usage conditions, it may not be possible 
to reach the indicated lowest temperature.
Caution: SANYO guarantees the product under certain warranty condi-
tions. SANYO in no way shall be responsible for any loss of content or 
damage to content.
ACCESSIBILITY
• A beak style inner 
door latch tightly 
closes the inner door 
against the freezer 
frame. It also helps make opening and 
closing the door smoother.
• The condenser filter is situated at the 
bottom right side of the front panel to 
make filter removal and cleaning easier.
RELIABILITY
• Status Alert (Condition monitor)
Monitors ambient and system conditions 
continuously and notifies of any 
abnormalities before a problem happens.
EVOLUTIONARY DESIGN
• SANYO V.I.P.™ Evolution Series -86˚C 
ultra-low temperature freezers use 
patented revolutionary vacuum insula-
tion cabinet construction that reduces 
the wall thickness from 14cm to 7cm 
(5.5” to 2.7”) and achieves up to 30% 
more storage capacity than a conven-
tionally insulated freezer without increas-
ing the footprint.
• The newly developed refrigeration 
system and the freezer structure offer a 
quiet experimental environment.
• Achieves 12% higher energy conser-
vation than previous model. (115V, 60Hz)
• The newly developed back cover com-
bined with new aerodynamically design-
ed and placed components in the 
refrigeration compartment provide 
superior air flow, making it possible to 
drastically reduce the stress on the 
freezer and contributing to excellent 
durability.
• Two independent and insulated inner 
doors with gaskets prevent cold air 
leakage.
Back 
cover
Compressors
Front
Rear
Air
5
25
20
15
10
30
35
10
15
20
5
T
ime (hours)
Ambient temperature (˚C)
T
e
mperature (˚C)
Time (hours)
40
20
0
–20
–120
–40
–140
–60
–160
–80
–100
0
12
6
5
4
3
2
10
1
11
9
8
7
Central air temp. in cabinet
Ambient temp. 30˚C, No-load
T
e
mperature (˚C)
Time (hours)
40
20
0
–20
–120
–40
–140
–60
–160
–80
–100
0
12
6
5
4
3
2
10
1
11
9
8
7
Ambient temp. and maintenance 
of –70˚C by injection of LCO
2
 (30kg)
Strip chart recorder
MTR-85H
Recorder mounting 
bracket 
MDF-S3085
(for MTR-85H)
105 (4.1)
30 
(1.2)
1639 (64.5)
18 (0.7)
490 (19.3)
670 (26.4)
450 (17.7)
1860 (73.2)
1410 (55.5)
45
(1.8)
733 (28.9)
1140 (44.9)
867 (34.1)
795 (31.3)
897 (35.3)
600 (23.6)
SANYO Electric Co., Ltd.
Biomedical Division
http://www.sanyo.co.jp/cmg/biomedical
SANYO Electric Co., Ltd., Biomedical Division, Gunma is certified for: 
Quality management system: ISO9001 / Medical devices quality management system: ISO13485 / Environmental management system: ISO14001
RoHS (European Restriction of Hazardous Substances directives) compliant
•Appearance and specifications are subject to change without notice.