Справочник Пользователя для National Instruments 321645c-01
Chapter 2
Function Reference — GPCTR_Set_Application
NI-DAQ FRM for PC Compatibles
2-224
©
National Instruments Corporation
Figure 2-17. Single Triggered Pulse Width Generation Measurement
Use the
GPCTR_Watch
function with entityID =
ND_COUNT_AVAILABLE
to monitor the
progress of the counting process. This measurement completes when entityValue becomes
ND_YES
. After this is completed, you can retrieve the counted value by using
GPCTR_Watch
with entityID =
ND_COUNT
, as shown in the following example code:
Create u32 variable count_available.
Create u32 variable counted_value.
repeat
{
GPCTR_Watch(deviceNumber, gpctrNumber, ND_COUNT_AVAILABLE,
count_available)
}
until (count_available = ND_YES)
GPCTR_Watch(deviceNumber, gpctrNumber, ND_COUNT, counted_value)
To calculate the measured interval, multiply the counted value by the period corresponding to
the timebase you are using. For example, if your
the timebase you are using. For example, if your
ND_SOURCE
is
ND_INTERNAL_20_MHZ
, the
interval will be 1/(20 MHz) = 50 ns. If the
ND_COUNT
is 4 (Figure 2-15), the actual interval is
4 * 50 ns = 200 ns.
Note
The measured interval will correspond to the most recent pulse that arrived prior
to the invoking of
to the invoking of
GPCTR_Watch
call with entityID set to
ND_COUNT_AVAILABLE
.
Caution
There should be source transitions between gate transitions in order for this
measurement to be correct.
measurement to be correct.
0
0
NO
NO
0
Gate
Source
Count
Count
Available
Available
Measured
Interval
NO
YES
0
0
1
2
2
Trigger
!