Chapter 2
Function Reference — SCAN_to_Disk
NI-DAQ FRM for PC Compatibles
2-318
©
 National Instruments Corporation
when you enable pretrigger mode. If you do not apply the stop trigger, 
SCAN_to_Disk
 
eventually returns control to your application because you eventually run out of disk space.
In any case, you can use 
Timeout_Config
 to establish a maximum length of time for 
SCAN_to_Disk
 to execute.
Note
Simultaneous sampling devices do not use the sampleRate parameter. Because 
these devices use simultaneous sampling of all channels the
 scanRate parameter 
controls the acquisition rate; therefore, a 
scanRate of 0 is not allowed.