source indicates what signal controls the track/hold state of the module. If the inputMode 
is 0, NI-DAQ ignores this parameter.
0:
A counter of the DAQ device that is cabled to the module will be the source 
(NI-DAQ will reserve and use Am9513-based device counter 2, an E Series 
dedicated DAQ-STC counter, Lab and 1200 Series devices counter B1, 
DAQCard-700 or LPM device counter 2 for this purpose). This source is only 
valid if the module is cabled to a DAQ device.
1:
An external signal connected to the HOLDTRIG pin on the front connector of 
the module will control the track/hold state of the module. There is a hardware 
connection between the HOLDTRIG pin and the counter output of the DAQ 
device, so if source = 1 the appropriate counter (listed above) is driven by the 
external signal and will be reserved. Keep in mind that if inputMode = 2, this 
external signal will drive the scan interval timer. If you are using a Lab and 1200 
Series devices, DAQCard-700, or LPM device, you must change the jumper 
setting on the SCXI-1341 or SCXI-1342 adapter device to prevent the external 
signal from damaging the timer chip on the DAQ device.
2:
NI-DAQ will use a signal routed on an SCXIbus trigger line to control the 
track/hold state of the module. If you are using an SCXI-1200 or the internal 
connection to the SCXI backplane on the PXI-1010, you must use this option to 
route the trigger signal from the backplane.
send indicates where else to send the signal specified by source for synchronization purposes. 
NI-DAQ also ignores this parameter if the inputMode is 0. 
0: Nowhere.
1: Make 
the 
source signal drive the DAQ device counter output and the 
HOLDTRIG pin on the module front connector (if the source is not already one 
of those signals). If you are using a DAQCard-700, DAQCard-1200, 
Lab-PC-1200, Lab-PC-1200AI, Lab-PC+, PCI-1200, or LPM device, you must 
change the jumper setting on the SCXI-1341 or SCXI-1342 adapter device to 
prevent the external signal from damaging the timer chip on the DAQ device.
2: Make 
the 
source signal drive an SCXIbus trigger line so that other SCXI-1140 
modules can use it (if the source is not from the SCXIbus). Only one SCXI-1140 
module can drive that trigger line; an error will occur if you attempt to configure 
more than one SCXI-1140 to drive it.
holdCount is the number of times the module is enabled by NI-DAQ during an interval scan 
before going back into track mode. Each time Slot 0 encounters an entry for the module in the 
module scan list, NI-DAQ enables the module, which remains enabled until the sample count 
in that module scan list entry expires. If there is only one entry for the module in the module 
scan list, holdCount should be 1 (this will almost always be the case).
Range:
1 to 255.