Chapter 2
Function Reference — AI_VRead_Scan
©
 National Instruments Corporation
2-19
NI-DAQ FRM for PC Compatibles
AI_VRead_Scan
Format
status = AI_VRead_Scan (deviceNumber, reading)
Purpose
Returns readings in volts for all analog input channels selected by 
SCAN_Setup
 (E Series 
devices only with or without the SC-2040 accessory).
Parameters
Input
 
Output
 
Parameter Discussion
reading is an array of readings from each sampled analog input channel. The length of the 
reading array is equal to the number of channels selected in the 
SCAN_Setup
 numChans 
parameter. NI-DAQ uses values you have specified in 
SCAN_Setup
 through the gains 
parameter for computing voltages. If you have attached an SC-2040 or SC-2043-SG to your 
DAQ device, NI-DAQ also uses values you have specified in 
SC_2040_Config
 (through 
the sc2040gain parameter) or 
Set_DAQ_Device_Info
 (a fixed gain of 10) for computing 
voltages.
If you have SCC modules connected, NI-DAQ also uses the SCC module gain for computing 
voltages.
Using This Function
AI_VRead_Scan
 samples the analog input channels selected by 
SCAN_Setup
, at half 
the maximum rate your DAQ hardware permits.You must use the 
SCAN_Setup
 function prior 
to invoking this function. 
You cannot use external signals to control A/D conversion timing and use this function at the 
same time.
Name
Type
Description
deviceNumber
i16
assigned by configuration utility
Name
Type
Description
reading
[f64]
voltage readings from each sampled analog 
input channel