Connect the power cord.
2
Plug the free end of the power cord into a power outlet. 
Connect the antibacterial filter firmly onto the air outlet of the device. 
4
Connect the external humidifier to the other side of the antibacterial filter.
 
*If not using an external humidifier, the HMEF can be connected to the leak port (at step 9). 
5
Connect the air tubing to the external humidifier.
6
Connect the leak port to the air tubing.
7
Select the mask type Trach (select Setup menu, then Clinical Settings, then Advanced 
Settings
).
8
9
If an external humidifier is not being used, you can connect the HMEF to the leak port if 
necessary.
10
Connect the catheter mount.
** The leak port or the HMEF can be connected to standardised tracheostomy interfaces 
including connector pieces like catheter mounts. 
The catheter mount and the external humidifier are not part of the ResMed component.
Antibacterial filter
External humidifier*
Air tubing
ResMed leak port
Catheter mount**
1
2
3
4
5
HMEF*
6
10
AC locking clip 
AC power cord
Leak Circuit 
configuration