4.5  Zero Calibration
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
NGA 2000
4 - 41
Measure
Cancel !
Back...
More...
TAG
F1
F2
F3
F4
F5
0.00 ppm
-- Calibration Procedure Status --
Procedure status:
Ready
Current/expected gas flow:
Samplegas
Concentration in span gas units:
0.00  ppm
Last zero calibration:
Success
Last span calibration:
Success
Last span calibration was:
Fri 05-29-1998 13:37:23
Successful zero+span calibrated ranges:
1+2+3+4
Remaining procedure time:
0  s
Last zero calibration was:
Mon 06-08-1998 09:21:34
---------------------- Results ----------------------
Display
Status...
Main...
Channel
Calib...
TAG
F1
F2
F3
F4
F5
Failures:
No
Maintenance-Requests:
No
Temperature:
20.0 C
Operation:
Ready
0.0
100.0
Range: 1
0.00
ppm CH4
0.00
50.00
⇒ 
⇒ 
Change to the
single component
display of the
current channel
Press  
F1
Options:
• 
F4 key: Changing back
to the menu "Analyzer
Module Calibration".
• 
F5 key: Changing to
the menu "Calibration
Deviations".
⇒ 
⇒ 
Single component
display appears after
zeroing is done