NI 5412 Specifications
14
ni.com
Delay from 
Start Trigger to 
CH 0 Analog 
Output
Interpolation Factor
Typical Delay
Refer to t
s2
 at 
NI Signal 
Generators 
Help»Devices»
NI 5412»
NI <bus>-5412»
Triggering»
Trigger Timing
.
Digital Interpolation Filter 
disabled.
43 Sample Clock 
Periods + 110 ns
2
57 Sample Clock 
Periods + 110 ns
4
63 Sample Clock 
Periods + 110 ns
8
64 Sample Clock 
Periods + 110 ns
Trigger Exporting
Exported 
Trigger 
Destinations
A signal used as a trigger can be routed out to any 
destination listed in the Destinations specification 
of Table 8.
Exported 
Trigger Delay
65 ns (typical). Refer to t
s3
 at NI Signal Generators Help»
Devices»NI 5412»NI <bus>-5412»Triggering»Trigger 
Timing
.
Exported 
Trigger Pulse 
Width
>150 ns. Refer to t
s4
 at NI Signal Generators Help»
Devices»NI 5412»NI <bus>-5412»Triggering»Trigger 
Timing
.
Table 7.   (Continued)