Справочник Пользователя для Siemens FX300

Скачать
Страница из 80
 OPERATION 
6
41
SITRANS FX300
www.siemens.com/flow
06/2009 - SFIDK.PS.050.F2.02   A5E02100423
6.2.3  Navigation within the menu structure
Navigation within the menu is by means of the  and ^ buttons. Pressing  button takes you one 
menu level lower, ^ takes you one menu level higher.
If you are already located at the lowest level (function level), you can use the  button to go in the 
change mode, which can be used to set data and values.
If you are located at the first level (main menu), you can use the ^ key to exit the menu mode and 
return to the measuring mode.
6.2.4  Changing the settings in the menu
•  Use the  and ↑ keys to navigate in the menu in which you want to change a setting or value. 
Use the  key to activate the change mode in the selected menu.
i
The current values or settings are displayed.
•  Use the  and ↑ keys to change the value or setting. 
Save the new value or setting using the ^ key.
Some menu items contain several setting options. They are displayed in sequence by pressing 
the ^ key.
i
Takes you back to the main menu.
•  Press the ^ key to save the settings.
i
The query "Save Yes" appears. Switch between "Yes" and "No" by pressing the ↑ key.
Switch between "YES" and "NO" by pressing the ↑ key.
Measuring 
mode
Main menu
Sub-menu
Function
Edit
→   
   
   
   ↑        ^
^
^
^
^
^
^
^
^
^
^
^
^
^
^
^
^
^
^
^
Table 6-3: Navigation menu structure
Save Yes
^
^
^
^
 Changes saved.
An update is carried out and the display returns to measuring mode.
Save No
^
^
^
^
 Changes not saved.
The display returns to measuring mode.
CAUTION!
Each time parameters or settings are changed, the measuring device carries out an internal 
plausibility check. 
If implausible inputs have been made, the display remains in the current menu, and the changes 
are not saved.
MA_Siemens_Sitrans_FX300_R02_en_PRT.book  Page 41  Monday, June 22, 2009  8:17 AM