1010N Dual Channel Temperature Sensor Inputs
3.2.3 NOTES ON SYSTEM 1010 ANALOG INPUT MODULES
Single Channel Models
z
System 1010 NEMA flowmeters report input module temperature sensor connections T1 and
T2 as system variables.
Dual Path Models
z
Dual Channel flowmeters report both T1 and T2 as generic variables for both channels.
z
The Dual Channel unit uses T1 for liquid temperature for Channel 1 and T2 as liquid tempera-
ture for Channel 2.
z
The Dual Path instrument uses T1 to report liquid temperature. (Note that T2 is used to report
ambient temperature in some Leak Detector systems.)
z
The Analog Input of temperature takes priority over the built-in RTD (Resistive Thermal Device)
measurement of temperature.
z
If you want both channels of a Dual-Channel meter to sense a single RTD, short TB1/2 to TB2/
2 and TB1/3 to TB2/3. This allows the same voltage to be fed to both channels. Once that is
complete, you must close the RTD current loop by connecting TB1/4+5 to TB2/4+5. (Note:
TB5 is just the cable shield.)