Dakota Ultrasonics 
18 
8)  Press the OK key to calculate the coating velocity and return to the menu 
screen, or ESC to cancel the one point calibration. 
9)  Finally, press the MEAS key to return to the measurement screen and begin 
taking readings. 
Note:  CHECK YOUR CALIBRATION!  Place the transducer back on the 
calibration point.  The coating  thickness reading should now match the known 
coating thickness sample.  If the thickness is not correct, repeat the steps 
above. 
 
2.7 Measure 
The CMX is now ready to measure.  There are two different measurement view 
options, each with a specific purpose – Digits & B-Scan.  The steps below outline 
how to toggle between the different view mode options: 
 
 
Selecting the Measurement View Option 
 
1)  Press the MENU key once to activate the menu items tab.  Press the MENU 
key multiple times to tab right and the ESC key multiple times to tab left until 
the DISP menu is highlighted and displaying the submenu items.