the job.  For example, if the measurement range was .080” (2.03mm) to .250” 
(6.35mm), the user would perform a one point calibration on a known thickness 
sample close to .250” (6.35mm), followed by a two point calibration close to .080” 
(2.03mm).  When a two point calibration is performed, the CMX calculates the zero 
and the velocity.  The following steps outline this procedure: 
 
Two Point Calibration 
 
1)  Physically measure an exact sample of the material or a location directly on 
the material to be measured using a set of calipers or a digital micrometer. 
2)  Apply a drop of couplant on the transducer and place the transducer in 
steady contact with the sample or actual test material.  Be sure that the 
reading is stable and the repeatability indicator, in the top left corner of the 
display, is fully lit and stable. Press the MENU key once to activate the 
menu items tab.  Press the MENU key multiple times to tab right and the 
ESC key multiple times to tab left until the CAL menu is highlighted and 
displaying the submenu items. 
3)  Use the UP and DOWN arrow keys to scroll through the sub menu items 
until MATL 2PT is highlighted.