16 Troubleshooting
English 202052A
Errors
Error Screen
Condition
Cause
Action
E-xx
E-1 through E-16, E-18
through E-99
Sensor not properly
connected to the sensor
cable.
Make sure the sensor is connected to the
sensor cable.
Sensor failure.
Replace sensor.
Cable failure.
• Replace sensor cable.
• Contact 3M Patient Warming technical
service.
E-xx
E-17
Sensor failure.
Replace sensor.
Cable failure.
• Replace sensor cable.
• Contact 3M Patient Warming technical
service.
Control unit failure.
Discontinue use of the unit. Contact a
biomedical technician.
E-xx
E-100 and up
Control unit failure.
Discontinue use of the unit. Contact a
biomedical technician.