Справочник Пользователя для Fluke 110
Models 110, 111 & 112
Users Manual
Users Manual
12
Accuracy ± ( [ % of Reading ] + [ Counts ] )
Function
Range
Resolution
Model 110
Model 111
Model 112
AC Volts
1,2
-- True RMS
(50 Hz to 500 Hz)
6000 mV
3
6.000 V
60.00 V
600.0 V
60.00 V
600.0 V
1 mV
0.001 V
0.01 V
0.1 V
0.001 V
0.01 V
0.1 V
1.0 % + 3
1.0 % + 3
1.0 % + 3
DC Volts
6000 mV
3
6.000 V
60.00 V
600.0 V
60.00 V
600.0 V
1 mV
0.001 V
0.01 V
0.1 V
0.001 V
0.01 V
0.1 V
0.7 % + 2
0.7 % + 2
0.7 % + 2
Continuity
600
Ω
1
Ω
Beeper guaranteed on < 20
Ω
, guaranteed off > 250
Ω
; detects
opens or shorts of 250
µ
s or longer.
Ohms
600.0
Ω
6.000 k
Ω
60.00 k
Ω
600.0 k
Ω
6.000 M
Ω
40.00 M
Ω
0.1
Ω
0.001 k
Ω
0.01 k
Ω
0.1 k
Ω
0.001 M
Ω
0.01 M
Ω
0.9 % + 2
0.9 % + 1
0.9 % + 1
0.9 % + 1
0.9 % + 1
1.5 % + 3
0.9 % + 1
0.9 % + 1
0.9 % + 1
0.9 % + 1
1.5 % + 3
0.9 % + 2
0.9 % + 1
0.9 % + 1
0.9 % + 1
0.9 % + 1
1.5 % + 3
0.9 % + 1
0.9 % + 1
0.9 % + 1
0.9 % + 1
1.5 % + 3
0.9 % + 2
0.9 % + 1
0.9 % + 1
0.9 % + 1
0.9 % + 1
1.5 % + 3
0.9 % + 1
0.9 % + 1
0.9 % + 1
0.9 % + 1
1.5 % + 3
Diode test
2.200 V
0.001 V
0.9 % + 2
1.9 % + 2
1.9 % + 2
1.9 % + 2
1.9 % + 2
1.9 % + 2
1.9 % + 2
1.9 % + 2
1.9 % + 2
1.9 % + 2
1.9 % + 2
1.9 % + 2
1.9 % + 2
1.9 % + 2
1.9 % + 2
1.9 % + 2
Capacitance
4
1000 nF
10.00
10.00
µ
F
100.0
µ
F
10000
µ
F
1 nF
0.01
0.01
µ
F
0.1
µ
F
1
µ
F
100
µ
F - 1000
µ
F: 1.9% + 2
> 1000
µ
F: 10% + 90 typical
AC Amps
5
-- True RMS
(50 Hz to 500 Hz)
(Models 111 and 112)
(Models 111 and 112)
10.00 A continuous or
20 A overload for 30
seconds maximum
20 A overload for 30
seconds maximum
0.01 A
NA
1.5 % + 3
1.5 % + 3