Справочник Пользователя для Agilent Technologies 6684A

Скачать
Страница из 156
 Verification   13
 
Programming The Tests
 
General Considerations
 Procedures are given for programming these tests either from the front panel keypad or from a GPIB controller. The
procedures assume you know how to use the front panel keypad or how to program over the GPIB (see the Power Supply
Operating Manual for more information). When using computer-controlled tests, you may have to consider the relatively
slow (compared to computer and system voltmeters) settling times and slew rates of the power supply. Suitable WAIT
statements can be inserted into the test program to give the power supply time to respond to the test commands.
 
 This power supply can provide more than 240VA at more than 2 volts. If the output connections touch each
other, severe arcing can occur resulting in burns, ignition or welding of parts. DO NOT ATTEMPT
 
           TO MAKE CONNECTIONS WHILE OUTPUT POWER IS ON. These connections should be performed
          
          only by qualified electronics personnel.
 
 Programming Parameters
 Table 2-2 lists the programming voltage and current values for each model. You can enter these values either from the front
panel or from a controller over the GPIB.
 
 Table 2-2.  Programming Voltage and Current Values
 
 Agilent Model
 Full Scale
 Max. Prog.
 Full Scale
 Max. Prog.
 Max. Prog.
 
 Voltage
 Voltage
 Current
 Current
 Overvoltage
 Agilent 6680A
 5V
 5.125V
 875A
 895A
 6.25V
 Agilent 6681A
 8V
 8.190V
 580A
 592A
 10.0V
 Agilent 6682A
 21V
 21.50V
 240A
 246A
 25.2V
 Agilent 6683A
 32V
 32.75V
 160A
 164A
 38.4V
 Agilent 6684A
 40V
 41.00V
 128A
 131A
 48.0V
 
 
General Measurement Techniques
 Figure 2-1 shows the setup for the Constant Voltage tests. Measure the dc output voltage directly at the sense (+S and -S)
terminals. Connect these terminals for remote sensing (to the +S and -S terminals). Connect these terminals for local
sensing. Be certain to use load leads of sufficient wire gauge to carry the output current (see Chapter 4 of the Power Supply
Operating Manual). To avoid noise pickup, use coaxial cable or shielded pairs for the test leads. If you use more than one
meter or a meter and an oscilloscope, connect separate leads for each instrument to avoid mutual-coupling effects.
 
 
Performance Test Record Sheets
 
 When performing the tests in this chapter, refer to the Performance Test Record sheets supplied at the end of this chapter.
Table 2-6 is for recording common information, such as, the test equipment used and the environmental conditions. Tables
2-7 through 2-11 are dedicated to specific models. Each sheet lists the acceptable test ranges for the model and provides a
place to record the results of the test.
 
 
 
Note
It is recommended that before you perform the tests in either Table 2-4 or Table 2-5, that you first locate
the appropriate Performance Test Record sheet from Tables 2-7 through Table 2-11 for your specific
model. Make a copy of this sheet, and record the actual observed values in it while performing the tests.
Use the sheets in Tables 2-7 through Table 2-11 as master reference sheets to run copies at any time.