Verification 19
Constant Current (CC) Tests
Test Setup
Connect the appropriate current monitoring resistor (see Table 2-l) as shown in Fig. 2-3. The accuracy of the resistor must
be as specified in the table.
Connect the appropriate current monitoring resistor (see Table 2-l) as shown in Fig. 2-3. The accuracy of the resistor must
be as specified in the table.
Test Procedures
The test procedures are given in Table 2-5. The tests are independent and may be performed in any order. The CC tests are:
• Current Programming and Readback Accuracy.
The test procedures are given in Table 2-5. The tests are independent and may be performed in any order. The CC tests are:
• Current Programming and Readback Accuracy.
• CC Load Effect.
• CC Source Effect.
• CC Noise (PARD).
Table 2-5. Constant Current (CC) Tests
Action
Normal Result
Current Programming and Readback Accuracy
This test verifies that the current programming and readback are within specification.
1
Turn off the power supply and connect the current monitoring resistor as
shown in Fig. 2-3. Connect a DVM across the resistor .
shown in Fig. 2-3. Connect a DVM across the resistor .
2
Turn on the power supply and program the output for 5 volts and 0
amperes.
amperes.
3
Short the load.
4
Observe the DVM voltage reading. Divide this by the resistance of the
current monitor resistor. Record the result as the Low Current value.
current monitor resistor. Record the result as the Low Current value.
Value within specified Low Current
limits.
limits.
5
Record the front panel display readback.
Value within specified readback limits.
6
Program output current to full scale (see Table 2-2).
7
Repeat Steps 4 and 5.
Both current readings within specified
High Current and readback limits.
High Current and readback limits.
CC Noise (PARD)
Periodic and random deviations (PARD) in the output (ripple and noise) combine to produce a residual ac current
superimposed on the dc output current. This test measures CC PARD, specified as the rms output current over the
frequency range of 20 Hz to 20 MHz.
superimposed on the dc output current. This test measures CC PARD, specified as the rms output current over the
frequency range of 20 Hz to 20 MHz.
1
Turn off the power supply and connect the current transformer, resistor,
capacitor and rms voltmeter (see Fig. 2-4).
capacitor and rms voltmeter (see Fig. 2-4).