3.0 INSTALLATION 
3.1 GENERAL 
INSTALLATION 
 
The scales should not be placed in a location that will reduce the accuracy.  
 
Avoid extremes of temperature. Do not place in direct sunlight or near air conditioning 
vents. 
 
Avoid unsuitable tables. The table or floor must be rigid and not vibrate.  
 
Avoid unstable power sources. Do not use near large users of electricity such as welding 
equipment or large motors. 
 
Do not place near vibrating machinery. 
 
Avoid high humidity that might cause condensation. Avoid direct contact with water. Do not 
spray or immerse the scales in water. 
 
Avoid air movement such as from fans or opening doors. Do not place near open windows 
or air-conditioning vents. 
 
Keep the scales clean. Do not stack material on the scales when they are not in use.  
 
3.2  PREPARATION FOR USE 
 
 
The scales come with a stainless steel platform, packed separately.  
 
Place the platform in the receptacles on the top cover. 
 
Do not press with excessive force as that could damage the load cell inside. 
 
If an optional scoop has been supplied, place it on top of the platform. 
 
If an optional display on Pillar is supplied, install the pillar in the upright position into the 
socket at the rear of the scale by carefully pushing the cable inside the pillar rod. Secure 
the pillar rod to the socket by using two M4 screws. A black plastic cap is provided to seal