5.  Turn the generator off. Check the stack for proper 
assembly, damaged components, or dirty mating 
surfaces. Recheck the ultrasound cable for proper 
connections. Repeat Steps 1 through 4. If the test still 
fails, go to the next step.
6.  Turn the generator off, remove the stack from the 
thruster, and remove the horn from the stack. Reinstall 
the transducer-booster assembly in the thruster and 
turn the generator on. Repeat Steps 2 through 4.
 
If the assembly passes with the horn removed, the horn 
is outside the specifications required for operation with 
the Digi-Trac preset range. Reassemble the horn to the 
stack. 
  The System tab of iQ Explorer has a section for 
Advanced Hardware. These features are designed for 
unique horns and applications. See Figure 5-4.
Consult Dukane Corporation before making adjustments 
to these settings.
Figure 5-4 
i
Q
 Explorer System Tab - Advanced Hardware
  If any failure indications are present with the horn 
removed, check the booster and the transducer for the 
following
•  Any visible damage
•  Loose or cracked stud
•  Pitted or dirty mating surfaces
Make any necessary repairs or adjustments.
Go to Step 7.
7.  Repeat Steps 1 through 6.
•  If the failure indications disappear, reassemble the 
horn to the stack and repeat Steps 2 through 4.
•  If any failure indications are still present, do not 
run this stack. Return the transducer and booster 
to Dukane for analysis.
Acoustic Stack Test
    Continued from previous page
For more information about stacks, see this 
article: www.dukane.com/us/SE_stackarticle.
NOTE
Page 76 
iQ  Series
 Ultrasonic Generator/Power Supply ES User’s Manual
Dukane Manual Part No. 403-575-00