ni.com
NI 9219 Timing Options
The NI 9219 supports four different timing options that are 
optimized for different types of applications by using different 
ADC conversion times. High Speed is optimized for high speed at 
the expense of noise rejection, Best 60 Hz Rejection is optimized 
for rejection of 60 Hz noise, Best 50 Hz Rejection is optimized for 
rejection of 50 Hz noise, and High Resolution is optimized for 
maximum overall noise rejection and provides good rejection of 
50 Hz and 60 Hz noise. Refer to the 
 section for more 
information.
NI 9219 Circuitry
The NI 9219 is channel-to-channel isolated. Four 24-bit 
analog-to-digital converters (ADCs) simultaneously sample all 
four analog input channels. An excitation circuit is enabled for 
all input modes that require excitation. The ADC and excitation 
circuits are reconfigured in each mode to accommodate each type 
of sensor. Refer to Figure 5 for an illustration of the input circuitry 
for one channel of the NI 9219.