© National Instruments Corporation
27
NI 9144 User Guide and Specifications
The calibration data is stored in a U32 array, though each offset field should 
be interpreted as a signed value.
Use the calibration coefficients with the following equation to generate 
corrected data:
Table 12.  NI 9203 Calibration Coefficients
Coefficient
Representation
Units
LSB Weight
Unsigned
pA/LSB
Offset
Signed
pA
Table 13.  NI 9203 Calibration Equation Information
Term
Units
Definition
I
corrected
pA
Calibrated current
I
raw
bits
The raw code from the 
NI 9203
I
expected0mA
bits
Expected code at 0 mA. 
0 bits for 0–20 mA range. 
32768 bits for ±20 mA 
range
LSB
weight
pA/bit
Number of pA in one bit
I
offset
pA
Offset at 0 mA
I
corrected
I
raw
(
)
I
raw
I
expected 0mA
∗ LSB
weight
I
offset
=
        pA                                bits                      pA/bit             pA
⎧ ⎪ ⎪ ⎨ ⎪ ⎪ ⎩
⎧ ⎪ ⎪ ⎪ ⎨ ⎪ ⎪ ⎪ ⎩
⎧ ⎪ ⎨ ⎪ ⎩
⎧ ⎨ ⎩