Recommended maximum cable length
1
PFI/CLKOUT, DC to 10 MHz ........200 m
CLKOUT High Gain, 105 MHz......80 m
2
PFI/CLKOUT Low Gain,
105 MHz..........................................30 m
3
Weight ....................................................0.459 lb (208 g)
Power Requirements
+3.3 V .....................................................800 mA, max
+12 V ......................................................700 mA, max
Environment
Maximum altitude...................................2,000 m (800 mbar) 
(at 25 °C ambient temperature)
Pollution Degree .....................................2
Indoor use only.
Caution
When required, clean the NI PXIe-6672 with a soft nonmetallic brush. Make sure 
that the device is completely dry and free from contaminants before returning it to service.
Operating Environment
Ambient temperature range ....................0 to 55 °C (Tested in accordance 
with IEC-60068-2-1 and 
IEC-60068-2-2. Meets 
MIL-PRF-28800F Class 3 
low temperature limit and 
MIL-PRF-28800F Class 2 
high temperature limit.)
Relative humidity range..........................10% to 90%, noncondensing 
(Tested in accordance with 
IEC-60068-2-56.)
1   
Cable length measurements were made with an RG 58 cable. Maximum cable length performance will vary depending on the 
cable type used.
2   
Maximum cable length with a direct cable connection. Loss from a signal splitter would reduce maximum cable length.
3   
Maximum cable length with a direct cable connection. Loss from a signal splitter would reduce maximum cable length.