developers, Measurement Studio offers a set of Visual C++ classes and 
tools to integrate those classes into Visual C++ applications. The libraries, 
ActiveX controls, and classes are available with Measurement Studio and 
NI-DAQ.
VI Logger is an easy-to-use yet flexible tool specifically designed for data 
logging applications. Using dialog windows, you can configure data 
logging tasks to easily acquire, log, view, and share your data. VI Logger 
does not require any programming; it is a stand-alone, configuration-based 
software.
Using LabVIEW, Measurement Studio, or VI Logger greatly reduces the 
development time for your data acquisition and control application.
Optional Equipment
NI offers a variety of products to use with the NI PCI-6110/6111, including 
the following cables, connector blocks, and other accessories:
Shielded cables and cable assemblies
Connector blocks, shielded 50- and 68-pin screw terminals
RTSI bus cables
For more specific information about these products, refer to the NI catalog 
at 
ni.com/catalog
.
Custom Cabling
NI offers cables and accessories for you to prototype your application or to 
use if you frequently change device interconnections.
When developing custom cabling, refer to the following guidelines:
For the AI signals, shielded twisted-pair wires for each AI pair yield 
the best results, assuming that you use differential inputs. Tie the shield 
for each signal pair to the ground reference at the source.
Route the analog lines separately from the digital lines.
When using a cable shield, use separate shields for the analog and 
digital halves of the cable. Failure to do so results in noise coupling 
into the analog signals from transient digital signals.
Mating connectors and a backshell kit for making custom 68-pin cables are 