Справочник Пользователя для National Instruments NI PCI-6110

Скачать
Страница из 118
Chapter 5
Calibration
5-2
ni.com
vary with time and temperature. It is better to self-calibrate when the device 
is installed in the environment in which it is used.
Self-Calibration
The NI PCI-6110/6111 can measure and correct for almost all of its 
calibration-related errors without any external signal connections. NI-DAQ 
provides a self-calibration method. This self-calibration process, which 
generally takes less than a minute, is the preferred method of assuring 
accuracy in your application. Initiate self-calibration to minimize the 
effects of offset, gain, and linearity drifts, particularly those due to warmup.
Immediately after self-calibration, the only significant residual calibration 
error could be gain error due to time or temperature drift of the onboard 
voltage reference. This error is addressed by external calibration, which is 
discussed in the following section. If you are interested primarily in relative 
measurements, you can ignore a small amount of gain error, and 
self-calibration should be sufficient.
External Calibration
The NI PCI-6110/6111 has an onboard calibration reference to ensure the 
accuracy of self-calibration. The specifications are listed in Appendix A, 
. The reference voltage is measured at the factory and stored 
in the EEPROM for subsequent self-calibrations. This voltage is stable 
enough for most applications, but if you are using the device at an extreme 
temperature or if the onboard reference has not been measured for a year or 
more, you may wish to externally calibrate the device. 
An external calibration refers to calibrating the device with a known 
external reference rather than relying on the onboard reference. 
Redetermining the value of the onboard reference is part of this process and 
the results can be saved in the EEPROM, so you should not have to perform 
an external calibration very often. You can externally calibrate the device 
by calling the NI-DAQ calibration function.
To externally calibrate the device, be sure to use a very accurate external 
reference. The reference should be several times more accurate than the 
device itself. For example, to calibrate a 16-bit device, the external 
reference should be at least ±0.001% (±10 ppm) accurate.
For a detailed calibration procedure for the NI PCI-6110/6111, click 
Manual Calibration Procedures at 
ni.com/calibration
.