NI 9871 Operating Instructions
Safety
Maximum Voltage
1
Connect only voltages that are within these limits.
RS485/RS422 Port-to-COM.............  –8 to +13 VDC max, 
Measurement Category I
V
SUP
-to-COM ...................................  ±28 V max,
Measurement Category I
Measurement Category I is for measurements performed on 
circuits not directly connected to the electrical distribution system 
referred to as MAINS voltage. MAINS is a hazardous live electrical 
supply system that powers equipment. This category is for 
measurements of voltages from specially protected secondary 
circuits. Such voltage measurements include signal levels, special 
equipment, limited-energy parts of equipment, circuits powered by 
regulated low-voltage sources, and electronics.
Caution
Do not connect to signals or use for 
measurements within Measurement Categories II, III, 
or IV.
1  
The maximum voltage that can be applied or output without creating a safety hazard.