Справочник Пользователя для National Instruments AT E Series

Скачать
Страница из 184
Chapter 4
Connecting Signals
© National Instruments Corporation
4-5
I/O Connector Signal Descriptions
Table 4-2. I/O Signal Summary for the AT E Series
Signal Name
Reference
Direction
Description
AIGND
Analog Input Ground—These pins are the reference point
for single-ended measurements and the bias current return
point for differential measurements. All three ground
references—AIGND, AOGND, and DGND—are connected
together on the AT E Series device.
ACH<0..15>
AIGND
Input
Analog Input Channels 0 through 15—Each channel pair,
ACH<i, i+8> (= 0..7), can be configured as either one
differential input or two single-ended inputs.
ACH<16..63>
AIGND
Input
Analog Input Channels 16 through 63 (AT-MIO-64E-3
only)—Each channel pair, ACH<ii+8> (= 16..23, 32..39,
48..55), can be configured as either one differential input or
two single-ended inputs.
AISENSE
AIGND
Input
Analog Input Sense—This pin serves as the reference node
for any of channels ACH <0..15> in NRSE configuration.
AISENSE2
AIGND
Input
Analog Input Sense (AT-MIO-64E-3 only)—This pin serves
as the reference node for any of channels ACH <16..63> in
NRSE configuration.
DAC0OUT
AOGND
Output
Analog Channel 0 Output—This pin supplies the voltage
output of analog output channel 0. This pin is not available
on the AT-AI-16XE-10.
DAC1OUT
AOGND
Output
Analog Channel 1 Output—This pin supplies the voltage
output of analog output channel 1. This pin is not available
on the AT-AI-16XE-10.
EXTREF
AOGND
Input
External Reference—This is the external reference input for
the analog output circuitry. This pin is not available on the
AT-MIO-16XE-10, AT-AI-16XE-10, or AT-MIO-16XE-50.
AOGND
Analog Output Ground—The analog output voltages are
referenced to this node. All three ground
references—AIGND, AOGND, and DGND—are connected
together on the AT E Series device.
DGND
Digital Ground—This pin supplies the reference for the
digital signals at the I/O connector as well as the +5 VDC
supply. All three ground references—AIGND, AOGND,
and DGND—are connected together on the AT E Series
device.
DIO<0..7>
DGND
Input or
Output
Digital I/O signals—DIO6 and 7 can control the up/down
signal of general-purpose counters 0 and 1, respectively.