Instruction Manual 
 
CON 400/410 
- 34 - 
7.6 P5.0 
 
Temperature 
7.6.1 
P5.1  Adjusting the temperature coefficient 
The temperature coefficient is the amount of change in conductivity per degree of temperature; 
it is expressed in percent per 
°C. Entering the exact temperature coefficient of your solution 
lets you accurately compensate temperature for almost any solution. You can adjust 0.0 to 
10.0 % per 
°C. Meter default is 2.1% per °C. Adjustment to 0.0% will not correct for 
temperature effects so the displayed value will be the actual value at the displayed 
temperature. 
 
From measurement mode 
1. Press 
SETUP key to enter Set Up mode.  
2. Press 
the 
MI/
S or MR/T keys to scroll through 
subgroups until you view parameter P5.0.  
3. Press 
the 
ENTER key to select parameter 5.1. The 
display shows “t.CO” on the upper display.  
4. Press 
the 
ENTER key again. The upper display 
shows the temperature coefficient and the lower 
display shows “t.CO”. 
5. Press 
the 
MI/
S or MR/T keys to select the 
temperature coefficient of your solution.  
6. Press 
the 
ENTER key to confirm selection and to 
proceed to step 3 of P5.2. Press the CAL/MEAS key 
twice to return to measurement mode.  
 
NOTES:  
If you do not know the temperature coefficient of your 
solution you can determine the correct value using the 
formula in Addendum 3 “Calculating Temperature 
Coefficients” on page 47.  
•   
Figure 30 - Changing the 
temperature coefficient