Instruction Manual
PC 10
- 9 -
6. Press the ENTER key to confirm temperature
calibration. The CON indicator will flash to
confirm calibration.
confirm calibration.
See Figure 13.
NOTES
To exit from Temperature Calibration mode without
confirming calibration, DO NOT press ENTER in step
confirming calibration, DO NOT press ENTER in step
6. Press CAL/MEAS instead.
Temperature calibration is restricted to ±5°C from
the initial value displayed during calibration (shown
in the secondary display).
the initial value displayed during calibration (shown
in the secondary display).
Since temperature readings affect the accuracy of the pH and conductivity measurements, it is
strongly recommended to carry out pH and conductivity calibration after a temperature
calibration is done.
strongly recommended to carry out pH and conductivity calibration after a temperature
calibration is done.