Instruction Manual 
 
Fluoride Electrode
 
 
20
TROUBLESHOOTING HINTS 
 
Symptom                 Possible Causes          
 
Next Step 
Out of Range            defective meter           
 
check meter with shorting strap 
Reading                                    
 
 
 
(see meter instruction manual) 
 
                         
defective electrode             
 
check electrode operation 
                  
                         
electrodes not plugged           
unplug electrodes and reset 
                        
in properly 
 
                         
reference electrode       
 
be sure reference electrode is 
                         
not filled              
 
 
filled 
 
                         
air bubble on membrane 
           remove bubble by re-dipping 
                         
 
 
 
 
 
electrode 
 
Noisy or Unstable 
defective meter           
 
check meter with shorting strap 
Readings (readings                            
continuously or 
air bubble on membrane 
 
remove bubble by re-dipping 
rapidly changing)        
 
 
 
 
electrode 
                   
    
 
            TISAB not used           
 
use recommended TISAB 
 
                         
meter or stirrer        
 
 
ground meter or stirrer 
                         
not grounded   
 
                         
defective electrode       
 
replace electrode 
 
                         
electrode exposed        
 
soak electrode in fluoride 
                         
to interference        
 
 
standard with added TISAB 
 
 
Drift (reading     
samples and standards   
 
allow solutions to come to room 
slowly changing        at different temperatures           
temperature before measurement 
in one direction)    
                         
electrode exposed to   
   
check section entitled 
                         
complexing agents                   
Complexation
 
                                              
                         
incorrect reference       
 
use recommended 
                         
filling solution       
 
 
filling solution 
 
Low Slope or        
standards contaminated 
     
prepare fresh standards 
No Slope          
or incorrectly made 
                       
                         
TISAB not used           
 
use recommended TISAB 
 
                         
standard used as TISAB          
use TISAB 
                       
                         
electrode exposed        
 
check section entitled