Auto-coding Function
1. IGM-1001D has automatic
code
code
recognition
function(e.g.
C26)
2. It recognizes the code number
automatically. This is a very
convenient function preventing
the inconvenience of setting the
code number on the glucose test
meter every time. Check the code
on the vial to see if the codes
match each other
automatically. This is a very
convenient function preventing
the inconvenience of setting the
code number on the glucose test
meter every time. Check the code
on the vial to see if the codes
match each other
CAUTION
1. Do not bend the glucose test strip to prevent the
automatic code recognition failure.
2. If the code recognition label is damaged, code
recognition may fail. Please check the code number on
the LCD window together with the code number on the
glucose test strip vial to see if code numbers match each
other.
the LCD window together with the code number on the
glucose test strip vial to see if code numbers match each
other.
Backlight Function
When
inserting
IGM-1001D
glucose
test
strip
(starting
measuring process), backlight
will turn on for 10 seconds.
After the test result displays,
backlight will last for 10 seconds.
By pressing the power button
(moving into memory mode),
backlight will turn on for 10
seconds.
When pressing the power button
for 3 seconds (entering the
setting mode), backlight will turn
on for 10 seconds.
When inserting the PC cable into
the USB port, backlight will turn
on for 10 seconds.
will turn on for 10 seconds.
After the test result displays,
backlight will last for 10 seconds.
By pressing the power button
(moving into memory mode),
backlight will turn on for 10
seconds.
When pressing the power button
for 3 seconds (entering the
setting mode), backlight will turn
on for 10 seconds.
When inserting the PC cable into
the USB port, backlight will turn
on for 10 seconds.
Performing a Control Solution Test
IGM-1001D control solutions contain a known amount of
glucose that reacts with IGM-1001D test strip. By comparing
your control solution test results with the expected range
printed on the test strip vial, it is able to check the meter and
test strips are working properly and that you are performing
the test correctly. It is very important that you perform the
check routinely to make sure you get accurate results.
The Glucose control solution should be used
Before using a new box of strips
Whenever you suspect the blood glucose test meter or
blood glucose test strip is not functioning properly.
If your blood glucose test results are not consistent with
your symptoms or if you think they are not accurate.
If you have dropped the blood glucose test meter.
For quality control in the point of care usage.
For teaching or learning the system.
When using the meter for blood glucose tests after any
For quality control in the point of care usage.
For teaching or learning the system.
When using the meter for blood glucose tests after any
disinfection procedure.
Firmly insert the test strip into the meter
test port. Insert down firmly in direction of
arrow on test strip (arrow up). Please do
not insert the glucose test strip upside
down.
When you insert the glucose test strip
into the glucose test meter, the power
automatically adjusts the meter
accordingly.
test port. Insert down firmly in direction of
arrow on test strip (arrow up). Please do
not insert the glucose test strip upside
down.
When you insert the glucose test strip
into the glucose test meter, the power
automatically adjusts the meter
accordingly.
Before inserting blood, Scroll through the
icons with the
icons with the
▲ or ▼ button until the
control solution bottle is blinking, press
1. Check the expiration date before
performing a control solution test.
Do not use if expired. Please note
the expiration date on the control
solution bottle.
Do not use if expired. Please note
the expiration date on the control
solution bottle.
2. Gently shake the control solution
before use.
3. Discard the 1
st
drop of control
solution, this will eliminate any
residue.
Place a drop of control solution on
a clean dry surface (e.g. the lid of
the test strip vial)
residue.
Place a drop of control solution on
a clean dry surface (e.g. the lid of
the test strip vial)
26