Справочник Пользователя для OSANG Healthcare Co. Ltd. IGM-1001D

Скачать
Страница из 16
   
INFOPIA CO., LTD                                                                                    Page 5                                                                                              UM-M (Rev.0) 
 
Auto-coding Function 
 
 
1.  IGM-1001D  has  automatic 
code 
recognition 
function(e.g. 
C26) 
 
 
 
 
 
 
 
 
2.  It  recognizes  the code  number 
automatically.  This  is  a  very 
convenient  function  preventing 
the  inconvenience  of  setting  the 
code  number  on  the  glucose  test 
meter every time. Check the code 
on  the  vial  to  see  if  the  codes 
match each other 
 
 
 
 
CAUTION 
1.  Do  not  bend  the  glucose  test  strip  to  prevent  the 
automatic code recognition failure. 
2.  If  the  code  recognition  label  is  damaged,  code 
recognition  may  fail.  Please  check  the  code  number  on 
the  LCD  window  together  with  the  code  number  on  the 
glucose test strip vial to see if code numbers match each 
other.   
 
 
 
Backlight Function 
 
 
 
When 
inserting 
IGM-1001D 
glucose 
test 
strip 
(starting 
measuring  process),  backlight 
will turn on for 10 seconds. 
 
After  the  test  result  displays, 
backlight will last for 10 seconds. 
   
By  pressing  the  power  button 
(moving  into  memory  mode), 
backlight  will  turn  on  for  10 
seconds. 
 
When  pressing  the  power  button 
for  3  seconds  (entering  the 
setting mode), backlight will turn 
on for 10 seconds. 
 
When inserting the PC cable into 
the USB port, backlight will turn 
on for 10 seconds. 
 
 
 
 
 
 
 
 
Performing a Control Solution Test 
 
IGM-1001D  control  solutions  contain  a  known  amount  of 
glucose that reacts with IGM-1001D test strip. By comparing 
your  control  solution  test  results  with  the  expected  range 
printed on the test strip vial, it is able to check the meter and 
test strips are working properly and that you are performing 
the  test  correctly.  It  is  very  important  that  you  perform  the 
check routinely to make sure you get accurate results. 
 
The Glucose control solution should be used 
  Before using a new box of strips 
  Whenever  you  suspect  the  blood  glucose  test  meter  or 
blood glucose test strip is not functioning properly. 
  If your blood glucose test results are not consistent with 
your symptoms or if you think they are not accurate. 
  If you have dropped the blood glucose test meter. 
  For quality control in the point of care usage. 
  For teaching or learning the system. 
  When using the meter for blood glucose  tests after any 
disinfection procedure. 
 
 
Firmly  insert  the  test  strip  into  the  meter 
test port. Insert down firmly in direction of 
arrow on test strip (arrow up). Please do 
not  insert  the  glucose  test  strip  upside 
down.   
When you insert the glucose test strip 
into the glucose test meter, the power 
automatically adjusts the meter 
accordingly. 
 
 
 
 
 
Before inserting blood, Scroll through the 
icons with the 
▲ or ▼ button until the 
control solution bottle is blinking, press 
 
 
   
 
1. Check  the  expiration  date  before 
performing  a  control  solution  test. 
Do  not  use  if  expired.  Please  note 
the  expiration  date  on  the  control 
solution bottle. 
 
 
 
2. Gently  shake  the  control  solution 
before use. 
 
 
 
 
 
 
3. Discard  the  1
st
  drop  of  control 
solution,  this  will  eliminate  any 
residue. 
Place a drop of control solution on 
a  clean  dry  surface  (e.g.  the  lid  of 
the test strip vial) 
 
26 
IGM-1001D