8 Solving problems
Solving problems
Solving problems
Solving problems
No. Symptom
Solution
1
The transmitter LED
does not flash after
you remove the
transmitter from the
charging cradle
does not flash after
you remove the
transmitter from the
charging cradle
•
Check that the transmitter is charged.
•
Re-insert the transmitter into a charging slot for a few
seconds and try again.
seconds and try again.
•
See Section 3.2.1 in case a reset of the device is required.
2
The receiver LED does
not start flashing
when the transmitter
LED starts flashing
not start flashing
when the transmitter
LED starts flashing
•
Check that the receiver is connected to its external power
supply.
supply.
•
Move the transmitter closer to the receiver.
•
Move the receiver away from other radio transmitting devices.
For example: laptops, mobile phones, etc.
For example: laptops, mobile phones, etc.
•
Reset the receiver by pulling out the power adapter. Plug in
the power adapter again after a few seconds.
the power adapter again after a few seconds.
3
You do not see a
signal in your
software.
signal in your
software.
•
Check that the transmitter is charged.
•
Make sure that external sensors are connected to the
transmitter.
transmitter.
•
Make sure that the EMG electrodes are attached to the
muscle.
muscle.
•
Check that the analog-to-digital device is correctly connected
to your computer.
to your computer.
•
Check that all the wires from the receiver’s analog output are
connected correctly.
connected correctly.
•
Check that pin 25 from the receiver’s analog output is
connected to ground.
connected to ground.
•
Reduce the distance between transmitters and receivers.
5
There are strong noise
/ hum artefacts on
EMG channels
/ hum artefacts on
EMG channels
•
Check that all interacting systems (PCs, notebooks, etc.) are
firmly connected to ground.
firmly connected to ground.
•
Make sure that you are using fresh electrodes from a recently
opened package and that their expiry date is not exceeded.
opened package and that their expiry date is not exceeded.