Справочник Пользователя для Fluke Electronics F381

Скачать
Страница из 48
 
Remote Display True-rms Clamp Meter 
 Measurements 
19 
Measurements 
Note 
Prior to first use, remove the battery isolator (small piece of plastic between the 
batteries and battery contacts). 
AC and DC Current (Jaw) 
 Warning 
To avoid electric shock or personal injury: 
 
When making current measurements, disconnect the test leads from the 
Meter. 
 
Keep fingers behind Tactile Barrier. See Figure 2 and Table 2. 
Note 
When measuring current, center the conductor in the Jaw using the alignment marks 
on the Jaw. 
Before taking dc measurements, push  to ensure correct readings. Zeroing the Meter removes 
dc offset from the reading. The Zero function works only in the dc current measurement Rotary 
Function Switch position.