381 
Users Manual 
22 
AC Current (Flexible Current Probe) 
 Warning 
To prevent possible electrical shock or personal injury: 
Do not apply the Flexible Current Probe around or remove from HAZARDOUS 
LIVE conductors. Take special care during fitting and removal of the Flexible 
Current Probe. De-energize the installation under test or wear suitable protective 
clothing. 
To use the Flexible Current Probe, follow these instructions: 
1. 
Connect the Flexible Current Probe to the Meter. See Figure 5.  
2. 
Connect the flexible part of the Flexible Current Probe around the conductor. If opening the 
end of the Flexible Current Probe to make the connection, make sure that you close and 
latch it. See the detail in Figure 5. You should be able to hear and feel the Flexible Current 
Probe lock snap into place. 
Note 
When measuring current, center the conductor in the Flexible Current Probe. If 
possible, avoid taking measurements close to other current-carrying conductors. 
3. 
Keep the probe coupling more than 2.5 cm (1 inch) away from the conductor.