2  Check the connection between DC power cable and SCU mini. 
3  Turn off the power switch and turn on again and then check the fan or status of back side. 
4  Replace another SCU and check the result of it 
5  Replace the fuse of standard SCU. (refer to Change fuse of SCU) 
6  Replace corresponding devices. 
8.1.3  Communication Test is failed 
 
Symptom 
 
Transmission error is occurred, failure of communication test is occurred. 
 
Expected Causes 
 
Network connection problem 
 
Network setting problem 
 
PC environment setting problem 
 
Wireless environment environment problem 
 
Devices failure 
 
Solutions 
1  Check the connection of network cable between Workstation and SCU. 
2  Check if the accurate network cable is used or not. (CAT 5E or 6) 
3  Set the network information of Workstation, SCU and detector again. 
4  Check whole workstation environment again such as firewall setting and release the power save mode. 
5  Check surrounding wireless communication environment. 
6  Boot up detector and SCU again by processing initialization. (Refer to <7.4 Product Initialization>.) 
7  Replace other SCU and Detectors and check the result of it 
8  Replace corresponding devices. 
8.1.4  The Active LED and Data LED of the Detector are blinking 
 
Symtom 
 
The active LED and data LED is blinking when power LED is ON. 
 
Expected Causes 
 
Detector registeration error 
 
Data transmission error 
 
Solutions 
1  Turn on SCU again 
2  Check the network cable connection 
3  Check the workstation environment and network information again 
4  Check if the surrounding wireless communication is good 
5  Check cable connection again when connected with tether interface cabel 
6  Replace other devices and check the result of it