Справочник Пользователя для Geophysical Survey Systems Inc. MINIHR

Скачать
Страница из 53
Quick Start Guide   
StructureScan Mini HR 
 
Geophysical Survey Systems, Inc. 
9
 
Scan Density: Selecting the Scan Density 
Highlight Scan Density. 
Press 
  
to toggle between High and Normal. 
• 
High – The Mini HR will scan at 8 scans/cm (approximately 240 scans/ft). Use this is when you are looking 
for very small objects. 
• 
Normal – The Mini HR will scan at 4 scans/cm (approximately 120 scans/ft). Use this is when you are 
looking for normal size objects. 
Note:  Disabled in 3D data collection mode 
Backlight: Setting the Brightness of the Mini HR Backlight 
Highlight Backlight. 
Press 
 
  to toggle between 25%, 50%, 75%, or 100%. 
Save Prompt: Whenever Saving, Have the System Prompt You to Save or Save Automatically 
Highlight the Save Prompt. 
Press 
  
 to toggle between On or Off. 
• 
On – The system will prompt you to save a file after you choose to save a file. 
• 
Off – The system will save the file automatically after you choose to save a file.