Sterilization
Sterilization
Caution!
Do not sterilize the ICS 3000!
If you are conducting pacemaker programming under 
sterile conditions, operate the programming head 
with a sterile cover (see „Optional Accessories“ on 
page 40).
Maintenance
The ICS 3000 requires no maintenance. The following 
inspections must be carried out:
Inspection A
Before each use, check the following:
Visually examine the ICS 3000 and the programming 
head.
Check the housing and cables for mechanical damage.
Check the system time and date and adjust if 
necessary.
Check if a replacement pen is available? (If not, order 
one.)
Check accessories, particularly the patient cables.
Inspection B
As part of the safety checks every year and if 
malfunctions are suspected:
First, conduct Inspection A.
Check all mechanical and electrical functions 
according to the BIOTRONIK test specification.
Check the accessories, particularly PK-222 and 
PK-199, and send them with the unit to the 
manufacturer if necessary.
Note:
The devices used for the electrical function test are 
usually not readily available in hospitals. Therefore, it 
is recommended that the ICS 3000 be checked by 
BIOTRONIK or by a test center authorized by 
BIOTRONIK.
Contact BIOTRONIK with any problems you may have: 
describe the problem, have program printouts or 
ECGs available, note error messages.