Power frequency 
(50/60Hz) mag-
netic field 
IEC 61000-4-8 
3.0 A/m
50 Hz or 60 Hz
3.0 A/m
50 Hz or 60 Hz
Power frequency mag-
netic fields should be 
at levels characteristic 
of a typical location in 
a typical commercial or 
hospital environment.
Voltage dips, short 
interruptions and 
voltage variations on 
power supply input 
lines IEC61000-4-11
0 % 
U
T
 ; 0,5 cycle
At 0° . 45°, 90°, 135°, 180°, 225°, 270°, and 315° 
q)
Mains power quaility 
should be that of a typical 
commercial or hospital 
environment.  If the user 
of the AMM240ED image 
intensifier requires con-
tinued operation during 
power mains interrup-
tions, it is recommended 
that the AMM240ED 
image intensifier be 
powered from an uninter-
ruptible power supply or 
a battery.
0% 
U
T
; 1 cycle 
and
70 % 
U
t
 ; 25/30 cycles
Single phase: at 0°
0 % 
U
t
 ; 250/300 cycle
Note: Ut is the A.C. mains voltage prior to application of the test level.
The Model AMM240ED is intended for use in the electromagnetic environment specified below. 
The customer or the user of AMM240ED should ensure that it is used in such an environment. 
Immunity Test
IEC 60601 Test Level
Compliance Level
Electromagnetic environment - 
guidance
Conducted RF
IEC61000-4-6
Radiated RF
IEC 61000-4-3
3 V
0,15 MHz - 80 MHz
6V in ISM and amature 
radio bands between 0,15 
MHz and 80 Mhz
80 % AM at 1 kHz
10 V/m
80 MHz - 2,7 GHz
80% AM at 1 kHz
3 V
0,15 MHz - 80 MHz
6V in ISM and amature 
radio bands between 
0,15 MHz and 80 Mhz
80 % AM at 1 kHz
10 V/m
80 MHz - 2,7 GHz
80% AM at 1 kHz
Portable and mobile RF communications 
equipment should be used no closer to 
any part of the 
AMM240ED, including 
cables, than the recommended separation 
distance calculated from the equation 
applicable to the frequency of the 
transmitter.
Recommended Separation Distance
0.15 MHz to 80 MHz
80 MHz to 800 MHz
800 MHz to 2.5 GHz
where 
P is the maximum oputput power 
rating of the transmitter in watts (
W
according to the transmitter manufacturer 
and d is the recommended separation 
distance in meters (
m).
Field strengths from fixed RF transmitters, 
as deter-mined by an electromagnetic 
site survey 
(a), should be less than the 
compliance level in each frequency range 
(b).Interferency may occur in the vicinity 
of equipment marked with the follwing 
symbol:
NOTE 1: At 80MHz and 800MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations.  Electromagnetic propagation is affected by absorption and
             reflection from structures, objects, and people.
(a) Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mo-
bile radios, amateur radio, AM and FM radio broadcast and TV broadcast, cannot be predicted theoretically with accu-
racy.  To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should 
be considered.  If the measured field strength in the location in which the AMM240ED is used exceeds the applicable 
RF compliance level above, the AMM240ED should be observed to verify normal operation. If abnormal performance is 
observed, additional measures may be necessary, such as reorienting or relocating the AMM240ED.
(b) Over the frequency range 150kHz to 80MHz, field strengths should be less than [V1] V/m.
Manufacturer’s Declaration - Electromagnetic Immunity
AMM240ED   