Справочник Пользователя для Renishaw plc RMP60Q
7.3
Fault-finding
Symptom
Cause
Action
Poor probe repeatability
and/or accuracy.
and/or accuracy.
Debris on part or stylus.
Clean part and stylus.
Poor tool change repeatability.
Redatum probe after each tool
change.
change.
Loose probe mounting on shank
or loose stylus.
or loose stylus.
Check and tighten as appropriate.
Calibration out of date and/or
incorrect offsets.
incorrect offsets.
Review probing software.
Calibration and probing speeds
not the same.
not the same.
Review probing software.
Calibration feature has moved.
Correct position.
Measurement occurs as stylus
leaves surface.
leaves surface.
Review probing software.
Measurement occurs within
the machine’s acceleration and
deceleration zone.
the machine’s acceleration and
deceleration zone.
Review probing software and
probe filter settings.
probe filter settings.
Probing speed too high or too
slow.
slow.
Perform simple repeatability trials
at various speeds.
at various speeds.
Temperature variation causes
machine and workpiece
movement.
machine and workpiece
movement.
Minimise temperature changes.
Machine tool faulty.
Perform health checks on
machine tool.
machine tool.
RMP60 status LEDs do
not correspond to RMI
or RMI-Q status LEDs.
not correspond to RMI
or RMI-Q status LEDs.
Radio link failure – RMP60 out of
RMI or RMI-Q range.
RMI or RMI-Q range.
Check position of RMI or RMI-Q,
see operating envelope.
see operating envelope.
RMP60 has been enclosed/
shielded by metal.
shielded by metal.
Change batteries.
RMP60 and RMI or RMI-Q are
not partnered.
not partnered.
Partner RMP60 and RMI or
RMI-Q.
RMI-Q.
Draft copy 09/07/12