Instructions for use
Implanting physicians should be experienced in epidural-access procedures and should be
thoroughly familiar with all product labeling.
w
Warning: DO NOT use the recharger on an unhealed wound. The recharging system
is not sterile, and contact with the wound can cause an infection.
#
Caution: If the neurostimulator is not being used for an extended period of time,
recommend that your patient charge the neurostimulator at least once per year. If the
battery is discharged, stimulation will stop and the neurostimulator may not
communicate with the controller.
#
Caution: Advise patients to charge the neurostimulator when a low battery message is
displayed on the controller in order to maintain uninterrupted therapy from the
neurostimulator. If the battery is discharged, stimulation will stop and the
neurostimulator may not communicate with the controller.
Note: The patient will be able to use the controller and recharger to charge a discharged
battery without causing damage to the battery or the neurostimulator.
#
Cautions:
When using sharp instruments near the neurostimulator, be extremely careful to
avoid nicking or damaging the case, the insulation, or the connector block.
Damaging the neurostimulator may require surgical replacement.
Do not use saline or other ionic fluids at connections, which could result in a short
circuit.
Charging the neurostimulator battery
Check the battery level of the neurostimulator before opening the package, and recharge the
neurostimulator if the battery is low. For charging instructions, refer to the recharging system
user manual. If the patient will be sent home with stimulation on, charge the neurostimulator
in the package before implant.
Verifying neurostimulator operation
Before opening the sterile neurostimulator package, verify that the neurostimulator is operable
by using the clinician programmer to interrogate the neurostimulator and read the
neurostimulator battery charge level. (Refer to the programming guide for instructions on how
to read the battery charge level.)
#
Caution: Do not implant a neurostimulator if it was dropped onto a hard surface from a
height of 30 cm (12 in) or more, because the neurostimulator may be damaged and fail
to operate properly.
Note: The neurostimulator pocket may be flushed with an antibiotic solution; do not submerge
the neurostimulator in fluid.
97715 2012-06  English 9