Справочник Пользователя для Given Imaging Limited CAPSH-3
PillCam Capsule Endoscopy
208
Printer
Printer
CD/DVD
Sensor Array
Cannot locate findings
•
Findings were not saved
under patient’s folder
under patient’s folder
•
Findings were saved with
the wrong name
the wrong name
•
•
Contact Customer Support
Study cannot be selected
in Study Manager
in Study Manager
Changing an archive location
that appears in the Study
Manager screen, the archive
remains on the archives list
but its study cannot be
selected
that appears in the Study
Manager screen, the archive
remains on the archives list
but its study cannot be
selected
Modify the archive location
manually
manually
Cannot open a video
from an archive in Study
Manager and an error
message requests user
to click the refresh button
from an archive in Study
Manager and an error
message requests user
to click the refresh button
Archive privileges set with no
access to read or write
access to read or write
Allow Read and Write access to
the archive folder
the archive folder
Problem
Cause
Action
Cannot print report
Printer is turned off
Turn printer on
Printer is not set as default
printer
printer
Set printer to Default Printer
Printer has a malfunction
Contact Customer Support
Problem
Cause
Action
Cannot burn CD/DVD
CD/DVD is not blank or
compatible with CD/DVD
ROM
compatible with CD/DVD
ROM
Contact Customer Support
Wrong burning procedure
Cannot eject CD/DVD
A video on the disc is open
Close the video and retry
Problem
Cause
Action
Connector is damaged
•
Mishandling
•
End of Life
Contact Customer Support
Sensor is torn from its
wire
Insulation of the sensor
wire is damaged
wire
Insulation of the sensor
wire is damaged
Problem
Cause
Action