Chapter 6
Maintenance
2100.4400.32
- 6.1 -
E-2
1
Maintenance Information
1.1
Overview of Maintenance Work
The device has been designed to provide years of operation with little maintenance. Main-
tenance of the hardware is limited to:
Software updates (as required) 
(for improved performance and for retrofitting options)
Replacement of fan after approx. 40000 operating hours
Replacement of fan mat (option)
Further actions:
Backup of parameter data to an external medium (recommended)
Adjustment of I/Q modulator (if required)
1.2
Required Tools
Software update
– PC
– Update data (on CD-ROM)
– RJ-45 cable (crossover or patch cable)
Backup of parameter data
– PC as backup medium
Replacement of fans
– Torx screwdriver TX10
Adjusting the I/Q Modulator
– Spectrum analyzer