Справочник Пользователя для Braemar Manufacturing LLC FUSIONMCT

Скачать
Страница из 17
 
 
 
 
      
    Fusion Wireless Recorder 
 
23 
Specifications 
Functional 
 
 
Fusion 
1, 2, or 3 channel 
Sample rate 
256 samples per second 
User interface 
LCD display and sound 
 
 
Memory 
 
Max total record time 
 
 
One channel 
30 days 
 
Two channel  
30 days 
 
Three channel 
20 days 
Type 
Flash 
Data retention 
Non-volatile 
 
 
Physical 
 
Dimensions 
4.1"x 2.25"x .75" 
 
Weight with batteries 
5.5 oz. 
Enclosure 
Molded plastic 
Operating position 
Any orientation 
 
 
Electrical 
 
Input impedance 
10M min. 
CMR ratio 
60dB 
AC signal range 
+/- 5mV 
DC signal range 
+/- 300mV 
Resolution 
12 bits 
Frequency response 
.05Hz to 80Hz 
FCC ID 
HHMFUSIONMCT 
IC ID 
9158A-FUSION 
 
 
Environmental 
 
Operating temperature 
0
°C to +45°C 
Non-operating temperature 
-20
°C to +65°C 
Operating humidity 
10% to 95% without condensation 
Non-Operating humidity 
5% to 95% without condensation 
 
 
Battery 
 
Type 
(2) AA Lithium Thionyl 
Life 
Battery life varies greatly due to local wireless 
coverage, distance to the wireless towers, Bluetooth 
usage, number of arrhythmia detections and other 
factors.  Typical life is approximately 8-10 days. 
 
 
Remove batteries during storage 
 
For optimum shelf life store batteries at ambient 
room temperature and 30% to 50% relative humidity 
 
 
Warranty 
12 months from shipment 
 
 
 
 
 
      
    Fusion Wireless Recorder 
 
24 
Electromagnetic Emissions 
Emissions test  Compliance 
Electromagnetic  
environment - guidance 
RF emissions 
CISPR 11 
Class B 
Fusion is suitable for use in all 
establishments, including domestic 
establishments and those directly 
connected to the public low-
voltage power supply network that 
supplies buildings used for 
domestic purposes. 
 
Electromagnetic Immunity 
Immunity test 
IEC 60601 
test level 
Compliance 
level 
Electromagnetic 
environment - guidance 
Electrostatic 
discharge (ESD)  
IEC 61000-4-2 
±6 kV contact 
±8 kV air 
±6 kV contact 
±8 kV air 
Floors should be wood, 
concrete or ceramic tile. If 
floors are covered with 
synthetic material, the relative 
humidity should be at least 
30% without condensation. 
Power frequency 
(50/60 Hz) 
magnetic field 
IEC 61000-4-8 
3 A/m 
3 A/m 
Power frequency magnetic 
fields should be at levels 
characteristic of a typical 
location in a typical 
commercial or hospital 