Advanced operation (Key-operator mode)
D
RYSTAR
 5500/5503 R
EFERENCE
 
MANUAL
Establishing the reference values and verifying image quality for 
general radiography applications (DT 2 B and DT 2 C)
After installation of a new Drystar 5500 and before initial use you must
establish Quality Control aim values. These values will be used as the base
line for comparison when daily Quality Control is done. These values should
be determined again after major service, repair or software update.
The following Quality Control aim values must be determined:
The daily operating density levels. Refer to 
Drystar 5500 image geometry. Refer to 
page
.
Once Quality Control aim values are established you must evaluate the
Spatial Resolution, the Artifact Levels and the Low Contrast Visibility to
determine if the image quality is acceptable. Refer to
 
 on page
.
The Quality Control aim values, the Spatial Resolution and Artifact Levels and
the Image Geometry values are all recorded on the Quality Control charts.
Refer to 
.
On these charts, the following test conditions are also recorded:
The type and serial number of the Drystar 5500.
The type and emulsion number of the film used to determine the reference 
values.
The type of densitometer used.
The time (day, month, year) that the values were established.
Before you can establish the daily operating levels, the Drystar 5500 must 
be switched on for at least 15 minutes and it must be calibrated as well. 
Refer to 
 on page