Remarks for HF-emission and immunity
D
RYSTAR
 5500/5503 R
EFERENCE
 
MANUAL
This device is intended for operation in an electromagnetic 
environment in which the radiated high frequency disturbance 
variables are monitored. The user of the device can help to 
prevent electromagnetic disruptions by maintaining the 
minimum distances between portable and mobile high frequency 
communication equipment (transmitters) and the device as 
recommended below, in accordance with the maximum output 
power of the communications equipment. 
Recommended Protective Distances between Portable and Mobile
High Frequency Communication Equipment and the Device
Rated Power of 
the Transmitter
W
Protective Distance in accordance with Transmission Frequency
m
150 kHz to 80 MHz
d = 1.2 
80 MHz to 800 MHz
d = 1.2 
800 MHz to 2.5 GHz
d = 2.3 
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
The distance can be determined through the equation for each respective column.
P is the rated power of the transmitter in watts (W) according to the manufacturer information on 
the transmitter, only for transmitters where the rated power is not mentioned in the above table.
• REMARK 1 : An additional factor of 10/3 has been used to calculate the recommended protective 
distance of transmitters in the frequency range from 80 MHz to 2.5 GHz, to reduce the probability 
that mobile portable communication equipment unintentionally brought into the area of the 
patients will lead to a disruption.
• REMARK 2 : These Guidelines may not be relevant in all situations. The dispersion of 
electromagnetic waves is influenced by absorption and reflections from buildings, objects and 
people.