the PS Receiver. 
4  Install the Mounting Plate.
Referring to the facility floor plan, decide the position of the 
PS Receiver. Do not place near metal objects and structural 
frames. Ensure that the service light will be visible and that 
space is allowed for the antennas. Unless there is a conduit 
tray, run the cables at least 1 ft. (30 cm) up and away from 
the PS Receiver. 
Install  the  mounting  plate  using  one  of  the  following 
methods. 
A  Fasten to a suspended ceiling guy wire, using:
•  1 multi-function Caddy clip 
•  2 machine screws
•  2 hex nuts
•  2 lock washers, OR 
B  Install directly to a ceiling or a wall using:
•  2 wood screws
•  2 hollow wall anchors, OR
C  Install to a receptacle box, using:
•  2 machine screws
5  Install the PS Receiver to the Mounting Plate.
Plug the wired terminal block into the PS Receiver and 
install the PS Receiver onto the Mounting Plate, using the 
screw removed in Step 2. DO NOT place excess network 
and power supply wires inside the PS Receiver case.
6  Attach the two antennas. 
The antennas have a protective covering to prevent them 
from touching metal. For best coverage, ensure that they 
are at least 2 ft. (60 cm) from metal objects. Do not encase, 
wrap, or otherwise cover the antennas.
Installation is complete. Proceed to Final Steps.
Final Steps
Do not supply power to the PS Receiver or any 
other network device until all connections are 
complete.
Once all devices are installed, power up the PS Receiver 
and other devices. In the PS Ssytem Manager, run Auto 
Configure to install the new devices, and then configure 
them for operation.
The coverage area of the PS Receiver must also be checked.
Installation Considerations
•  Position the PS Receiver beneath ceiling obstruc-
tions—When mounting above drop ceilings, mount the 
PS Receiver below metal duct work, pipes and foil-backed 
ceiling tiles.
•  Route cables up and away from the PS Receiver—
When the cable is not in conduit, keep it at least 1 ft. 
(30 cm) from the PS Receiver. 
•  Install the PS Receiver temporarily until testing is 
complete—Fastenings should be secure enough that 
the PS Receiver can be permanently mounted without 
disturbing its orientation.
•  Point  the  secondary  antenna  into  a  clear  area, 
away from metal objects—The main antenna always 
points downwards, but the orientation of the second-
ary antenna is adjusted to suit site conditions. Point 
it away from metal objects, and make sure it is not 
running parallel to cabling in the ceiling. 
•  Observe temperature and humidity restrictions
Refer to the Specifications table. 
•  Leave slack in the cable—Leave at least 10 ft. (3 m) of 
slack in the communication and power cable to allow 
for adjustments in the location of the PS Receiver. 
•  No user adjustments to internal circuitry—Tamper-
ing may cause component or system failure, or both, 
and will void the warranty.
Installing the PS Receiver
• Ensure that the power supply and network 
are turned off to prevent electrical shock or 
damage to equipment. 
• Touch  your  hand  to  ground  to  discharge 
any electrostatic charge before handling the 
PS Receiver.
1  Record the LonWorks Neuron ID number.
Locate and record the LonWorks Neuron ID number onto 
the facility floor plan indicating the PS Receiver’s position 
and network wiring path. Affix one of the supplied Neuron 
ID labels on the floor plan and affix the other on the PS 
Receiver in a visible location.
2  Remove the Mounting Plate from the PS Receiver. 
If the PS Receiver is to be installed to a receptacle box, pass 
the network and power supply wires through the hole in 
the Mounting Plate. (For other installations, the wires are 
passed through the slot on the side of the PS Receiver case 
in Step 5.)
3  Wire the terminal block.
Connect the LonWorks Network IN, Network OUT, and 
Power Supply to the terminals. If the PS Receiver is at the end of 
the network, then terminate the network segment with Item A  
(105 Ω terminator). Ensure there is at least 10 ft. (3 m) 
of slack in the cable to allow the possible repositioning of 
Patient Security Receiver Installation Guide