2.  CHARACTERISTICS AND ENVIRONMENTAL CONDITIONS 
Usage conditions 
 
Temperature 
Stable between 0°C and 45°C 
Maximum dynamic acceleration measured 
-6g à 6g 
orientation measures are valid for static and quasi static positions 
 
The measured orientation parameter is valid in a uniform or quasi uniform magnetic field between  -6 
gauss and +6 gauss
1
 
 
Shock resistance 
4600 g for a time of 0,5 ms 
 MOTIONPOD™ should not be thrown away 
violently. It is however resistant to shocks 
for normal conditions of use. 
Magnetic field survivability 
10 000 Gauss 
Maximum rate of magnetic field tolerated before miss-
calibration  
20 Gauss 
 
 
 
 
Packaging and transport conditions
 
Temperature 
Between -20°C and 45°C (For 1 month 
maximum) 
Shock resistance 
4600 g for a time of 0,5 ms 
Magnetic field survivability 
10 000 Gauss 
Maximum rate of magnetic field tolerated before miss-
calibration 
20 Gauss 
 
 
Storage conditions 
 
Temperature 
Between -20°C à 35°C (For 6months 
maximum) 
45°C (For 1 month maximum) 
Shock resistance 
4600 g for a time of 0,5 ms 
Magnetic field survivability 
10 000 Gauss 
Maximum rate of magnetic field tolerated before miss-
calibration 
20 Gauss 
 
 
MOTIONPOD™ transmits data to the MOTIONCONTROLLER™ via radio signal. The 
transmitted signal is a Bluetooth band. MOTIONPOD™ is tested in CEM. However, 
band obstruction caused by the presence of other radio operating devices can lead 
to interferences with MOTIONPOD™. Do not to use the MOTIONPOD™ when other 
radio operating devices are running and are priority for patients. 
                                                             
1
 MotionDevTool v2.0.4 software automatically checks if this condition is verified during recording of the 
movement signature. An alarm is emitted in case of magnetic disturbance. If there is no alarm, the user is 
under desired conditions.